Teledyne LeCroy LabMaster 10Zi Getting Started Manual User Manual
Page 160
Getting Started Manual
LM10Zi-GSM-E Rev A
153
By selecting All, the test are true only if every waveform falls
within the set limit. Selecting Any makes the test true if just one
waveform falls within the limit.
2.
Touch inside the Condition field on the ParamCompare part of
the dialog and select a math operator from the pop-up menu.
3.
Touch inside the Limit field and enter a value using the pop-up
numeric keypad. The value entered acquires the dimensions
corresponding to the parameters being tested. For example, if
you are testing a time parameter, the unit is seconds.
If you choose
WithinDeltaPct or
WithinDeltaAbs from
the Condition menu, you also have the choice of setting the limit
by means of the statistical buttons (relating to standard statistical
limits) at the bottom of the ParamCompare dialog:
Comparing Dual Parameters
In Dual Parameter Compare mode, your oscilloscope can take
parameter results measured on two different waveforms and compare
them to each other. If desired, set your test to be true if Any or All
waveforms fit the criteria set by the comparison condition. Setups are
conveniently shown in the Summary box of the Qx dialog as follows
(shown on the Dual Param Compare right-hand dialog):
1.
Touch Analysis → Pass/Fail Setup... from the menu bar.
2.
Touch one of the Qx tabs and a setup dialog for the position is
shown.