3.
Touch inside the Condition field in the main dialog and select
DualParamCompare.
4.
Touch inside the Source1 and Source2 fields and select a source
from the pop-up menu.
5.
Touch inside the ParamCompare mini-dialog field and select a
source from the pop-up menu.
6.
Touch inside the Compare Values field and select All or Any from
the pop-up menu.
By selecting All, the test are true only if every waveform falls
within the set limit. Selecting Any makes the test true if just one
waveform falls within the limit.
7.
Touch inside the Condition field on the ParamCompare part of
the dialog and select a math operator from the pop-up menu.
8.
Touch inside the Limit field and enter a value, using the pop-up
numeric keypad. This value takes the dimension of the parameter
that you are testing. For example, if you are testing a time
parameter, the unit is seconds.
9.
If you choose either
WithinDeltaPct or
WithinDeltaAbs from the Condition menu respectively, touch
inside the Delta field and enter a value.
Also see Mask Testing (on page 158).