oscilloscope automatically sequences through all the trigger
setups.
PLEASE NOTE THE FOLLOWING:
You can tune the dwell time that the oscilloscope waits
before loading the next trigger setup using the Dwell Time
data entry control.
If you have Persistence display mode enabled, all trigger
events are recorded on the display. Refer to the
Persistence Setup
topic for instructions on
enabling Persistence display mode.
If you want TriggerScan to stop when the next oscilloscope
trigger, check the Stop On Trigger checkbox. You can use
this to isolate specific trigger setups.
Saving TriggerScan Setups
You should save TriggerScan setups once you have made any
modifications to the Trigger List. The current Trigger List is not be
preserved after exiting the application unless you save it.
1.
On the TriggerScan dialog, touch inside the Setup File Name data
entry control and providing a file name using your preferred
input control method. Additional information on using the touch
screen controls can be found in the Dialog Area.
OR
Touch the Browse button and select a location and file name.
2.
Touch the Save Setup... button.
Note: You can load previously saved TriggerScan setups by touching the
Browse button, locating the file, and then touching Load Setup....