beautypg.com

Teledyne LeCroy X-STREAM OSCILLOSCOPES Remote Control User Manual

Page 59

background image

C

H A P T E R

F

O U R

:

Understanding and Managing Waveforms

WM-RCM-E Rev D

ISSUED: February 2005

53

WAVEFORM?. The examination of data values for waveforms with two data arrays can be performed as
follows:

INSPECT? “DUAL”

to get pairs of data values on a single line

INSPECT? “DATA_ARRAY_1”

to get the values of the first data array

INSPECT? “DATA_ARRAY_2”

to get the values of the second data array

INSPECT? has its limitations; it is useful, but also wordy. INSPECT? cannot be used to send a waveform back

to the oscilloscope. If you want to do this, or you want the information quickly, you should instead use
WAVEFORM. With WAVEFORM_SETUP it is possible to examine just a part of the waveform or a sparsed

form of it. See the following pages.

If you’re a BASIC user you might also find it convenient to use INSPECT? and WAVEFORM? together to

construct files containing a version of the waveform descriptor that both you and BASIC can read. Using a
stored waveform, this can be done in a format suitable for retransfer to the instrument with MC:INSPECT?

“WAVEDESC”;WAVEFORM?, and then placing the response directly into a disk file.