beautypg.com

Remote control commands and queries, Waveform, Transfer – Teledyne LeCroy X-STREAM OSCILLOSCOPES Remote Control User Manual

Page 233

background image

Remote Control Commands and Queries

WM-RCM-E Rev D

ISSUED: February 2005

227

WAVEFORM

TRANSFER

STORE, STO

Command

DESCRIPTION

The STORE command stores the contents of the specified trace in
one of the internal memories M1 to M4 or in the current directory

in mass storage.

COMMAND SYNTAX

STOre

[,]

: = {F1,F2,F3,F4,F5,F6,F7,F8,TA,TB,TC,TD
,C1

,C2,C3, C4,ALL_DISPLAYED}. TA through TD are for

compatibility with existing software with earlier instruments. These

four mnemonics are not returned by queries.

: = {M1, M2, M3, M4, FILE}

TIP: If you send the

STORE

command without an argument,

all traces currently enabled in the Store Setup will be stored.

Modify this setup using

STORE_SETUP

.

AVAILABILITY

: = {C3, C4} only available on four-channel
oscilloscopes.

EXAMPLE (GPIB)

The following instruction stores the contents of Trace A (TA) into
Memory 1 (M1):

CMD$=“STO F1,M1”: CALL IBWRT(SCOPE%,CMD$)

The following instruction stores all currently displayed waveforms

onto the memory card:

CMD$=“STO ALL_DISPLAYED, FILE”:

CALL

IBWRT(SCOPE%,CMD$)

The following instruction executes the storage operation currently

defined in the Storage Setup (see command STORE_SETUP):

CMD$=“STO”: CALL IBWRT(SCOPE%,CMD$)

RELATED COMMANDS

STORE_SETUP