beautypg.com

Teledyne LeCroy X-STREAM OSCILLOSCOPES Remote Control User Manual

Page 132

background image

P

A R T

T

W O

:

C O M M A N D S

126

ISSUED: February 2005

WM-RCM-E Rev D

FUNCTION

DEFINE, DEF

Command/Query

DESCRIPTION

The DEFINE command specifies the mathematical expression to
be evaluated by a function. This command is used to control all

math tools and zoom in the standard oscilloscope as well as those in
the optional math software packages. See the Operator’s Manual

for details.

COMMAND SYNTAX

: DEFine EQN,‘
[,,,...]

N

N

O

O

T

T

E

E

:

: Function parameters are grouped in pairs. The first in the pair names the variable to be

modified, , while the second one gives the new value to be assigned. Pairs can be given

in any order and restricted to the variables to be changed.
Space (blank) characters inside equations are optional.

QUERY SYNTAX

: DEFine?

RESPONSE FORMAT

: DEFine EQN,‘’[,MAXPTS,]
[,SWEEPS,][,WEIGHT,][,BITS,]

F

UNCTION

P

ARAMETERS

Description

BITS

Number of ERES bits

CENTER

Horizontal

center

position for histogram display.

EQN

’ Function equation as defined below

LENGTH

Number of points to use from first waveform

MAX_EVENTS

Maximum

number of values in histogram

MAXBINS

Number of bins in histogram

START