ADDITIONAL INFORMATION
S
T A T U S
B
Y T E
R
E G I S T E R
(STB)
Bit
Bit Value
Bit Name
Description
Note
7
128
DIO7
0 reserved for future use
6 64
MSS/RQS
MSS = 1
RQS = 1
at least 1 bit in STB masked by SRE is 1
service is requested
1.
2.
5
32
ESB
1 an ESR enabled event has occurred
3.
4
16
MAV
1 output queue is not empty
4.
3 8
DIO3 0 reserved
2
4
VAB
1 a command data value has been adapted
5.
1 2
DIO1 0 reserved
0
1
INB
1 an enabled internal state change has occurred
6.
N
N
O
O
T
T
E
E
:
: For the above table...
1.
The Master Summary Status (MSS) indicates that the oscilloscope requests service, while the
Service Request status (when set) specifies that the oscilloscope issued a service request. Bit
position 6 depends on the polling method:
Bit 6 = MSS if an
*STB?
query is received,
= RQS if serial polling is conducted.
2.
Example: If SRE = 10 and STB = 10, then MSS = 1. If SRE = 010 and STB = 100 then MSS = 0.
3.
The Event Status Bit (ESB) indicates whether or not one or more of the enabled IEEE 488.2
events have occurred since the last reading or clearing of the Standard Event Status Register
(ESR). ESB is set if an enabled event becomes true (1).
4.
The Message AVailable bit (MAV) indicates whether or not the Output queue is empty. The MAV
summary bit is set true (1) whenever a data byte resides in the Output queue.
5.
The Value Adapted Bit (VAB) is set true (1) whenever a data value in a command has been
adapted to the nearest legal value. For instance, the VAB bit would be set if the timebase is
redefined as 2.5 µs/div since the adapted value is 2 µs/div.
6.
The INternal state Bit (INB) is set true (1) whenever certain enabled internal states are entered.
For further information, refer to the
INR