DISPLAY

PERSIST_SETUP, PESU

Command/Query

DESCRIPTION

The PERSIST_SETUP command selects the persistence duration
of the display, in seconds, in persistence mode. In addition, the

persistence can be set either to all traces or only the top two on the
screen.

The PERSIST_SETUP? query indicates the current status of the
persistence.

COMMAND SYNTAX

PErsist_SetUp

: = {PERTRACE, ALL}
Note: This does not support the argument “Top2” of earlier

instruments.

QUERY SYNTAX

PErsist_SetUp?

RESPONSE FORMAT

PErsist_SetUp

EXAMPLE (GPIB)

The following instruction sets the variable persistence at 10 seconds
on all traces:

CMD$=“PESU 20,ALL”: CALL IBWRT(SCOPE%,CMD$)

RELATED COMMANDS

PERSIST, PERSIST_COLOR, PERSIST_PERS,