Teledyne LeCroy X-STREAM OSCILLOSCOPES Remote Control User Manual

Page 148

background image

P

A R T

T

W O

:

C O M M A N D S

142

ISSUED: February 2005

WM-RCM-E Rev D

ADDITIONAL INFORMATION

S

TANDARD

E

VENT

S

TATUS

R

EGISTER

(ESR)

Bit

Bit Value Bit Name Description

See Note ...

15...8

0

Reserved by IEEE 488.2

7

128

PON

1

Power off-to-ON transition has occurred

1.

6 64 URQ

1

not

used

2.

5 32 CME

1

CoMmand

parser

Error has been detected

3.

4 16 EXE

1

EXecution

Error

detected

4.

3

8

DDE

1

Device Dependent (specific) Error occurred

5.

2

4

QYE

1

QuerY Error occurred

6.

1

2

RQC

0

Oscilloscope never ReQuests bus Control

7.

0 1 OPC

0

OPeration

Complete bit not used

8.

N

N

O

O

T

T

E

E

S

S

:

: (refer to table above)

1.

The Power On (PON) bit is always turned on (1) when the unit is powered up.

2.

On older LeCroy scopes, this bit reports softkey inputs. It does not apply to X-Stream scopes.

3.

The CoMmand parser Error bit (CME) is set true (1) whenever a command syntax error is

detected. The CME bit has an associated CoMmand parser Register (CMR) which specifies the

error code. Refer to the query CMR? for further details.

4.

The EXecution Error bit (EXE) is set true (1) when a command cannot be executed due to some

device condition (e.g., oscilloscope in local state) or a semantic error. The EXE bit has an

associated Execution Error Register (EXR) that specifies the error code. Refer to query EXR? for

further details.

5.

The Device specific Error (DDE) is set true (1) whenever a hardware failure has occurred at

power-up or at execution time, such as a channel overload condition, or a trigger or timebase

circuit defect. The origin of the failure can be localized with the DDR? query.