MISCELLANEOUS

*OPT?

Query

DESCRIPTION

The *OPT? query identifies oscilloscope options: installed software
or hardware that is additional to the standard instrument

configuration. The response consists of a series of response fields
listing all the installed options.

QUERY SYNTAX

*OPT?

RESPONSE FORMAT

*OPT

,,..,

: = A three- or four-character ASCII string

N

N

O

O

T

T

E

E

:

: If no option is present, the character 0 will be

returned.

EXAMPLE (GPIB)

The following instruction queries the installed options:

CMD$=“*OPT?”: CALL IBWRT(SCOPE%,CMD$):

CALL IBRD(SCOPE%,RSP$): PRINT RSP$

If, for example, the waveform processing options DFP2, SDM, JTA2,

and GPIB are installed, the response will be returned as:

*

DFP2,SDM,JTA2,GPIB

Response message if no options are installed:

*OPT

0