Remote Control Commands and Queries

WM-RCM-E Rev D

ISSUED: February 2005

143

6.

The QuerY Error bit (QYE) is set true (1) whenever (a) an attempt is made to read data from the

Output Queue when no output is either present or pending, (b) data in the Output Queue has

been lost, (c) both output and input buffers are full (deadlock state), (d) an attempt is made by the

controller to read before having sent an , (e) a command is received before the response to

the previous query was read (output buffer flushed).

7.

The ReQuest Control bit (RQC) is always false (0), as the oscilloscope has no GPIB controlling

capability.

8.

The OPeration Complete bit (OPC) is set true (1) whenever *OPC has been received, since

commands and queries are strictly executed in sequential order. The oscilloscope starts processing

a command only when the previous command has been entirely executed.