ACQUISITION

TIME_DIV, TDIV

Command/Query

DESCRIPTION

The TIME_DIV command modifies the timebase setting. The new
timebase setting can be specified with units: NS for nanoseconds,

US for microseconds, MS for milliseconds, S for seconds, or KS for
kiloseconds. Alternatively, you can use exponential notation: 10E-6,

for example. An out-of-range value causes the VAB bit (bit 2) in the
STB register (see table on page 225) to be set.

The TIME_DIV? query returns the current timebase setting.

COMMAND SYNTAX

Time_DIV

: = See the Operator’s Manual for specifications.

The unit S (seconds) is optional.

QUERY SYNTAX

Time_DIV?

RESPONSE FORMAT

Time_DIV

EXAMPLE (GPIB)

The following instruction sets the time base to 500 µs/div:

CMD$=“TDIV 500US”: CALL IBWRT(SCOPE%,CMD$)

The following instruction sets the time base to 2 msec/div:

CMD$=“TDIV 0.002”: CALL IBWRT(SCOPE%,CMD$)

RELATED COMMANDS