ACQUISITION

AUTO_SETUP, ASET

Command

DESCRIPTION

The AUTO_SETUP command attempts to display the input signal(s)
by adjusting the vertical, timebase and trigger parameters.
AUTO_SETUP

operates only on the channels whose traces are

currently turned on. If no traces are turned on, AUTO_SETUP
operates on all channels.

If signals are detected on several channels, the lowest numbered
channel with a signal determines the selection of the timebase and

trigger source.

If only one input channel is turned on, the timebase will be adjusted

for that channel.

The : AUTO_SETUP FIND command adjusts gain and
offset only for the specified channel.

COMMAND SYNTAX

: Auto_SETup [FIND]

: = {C1, C2, C3, C4}

If the FIND keyword is present, gain and offset adjustments will be
performed only on the specified channel. If no prefix is

added, an auto-setup will be performed on the channel used on the
last ASET FIND remote command. In the absence of the FIND
keyword, the normal auto-setup will be performed, regardless of the
prefix.

AVAILABILITY

: = {C3, C4} only on four-channel oscilloscopes.

EXAMPLE

The following instructs the oscilloscope to perform an auto-setup:

CMD$=“ASET”: CALL IBWRT(SCOPE%,CMD$)