Remote control commands and queries, Acquisition – Teledyne LeCroy X-STREAM OSCILLOSCOPES Remote Control User Manual
Page 101

Remote Control Commands and Queries
WM-RCM-E Rev D
ISSUED: February 2005
95
ACQUISITION
AUTO_SETUP, ASET
Command
DESCRIPTION
The AUTO_SETUP command attempts to display the input signal(s)
by adjusting the vertical, timebase and trigger parameters.
AUTO_SETUP
operates only on the channels whose traces are
currently turned on. If no traces are turned on, AUTO_SETUP
operates on all channels.
If signals are detected on several channels, the lowest numbered
channel with a signal determines the selection of the timebase and
trigger source.
If only one input channel is turned on, the timebase will be adjusted
for that channel.
The
offset only for the specified channel.
COMMAND SYNTAX
If the FIND keyword is present, gain and offset adjustments will be
performed only on the specified channel. If no
added, an auto-setup will be performed on the channel used on the
last ASET FIND remote command. In the absence of the FIND
keyword, the normal auto-setup will be performed, regardless of the
AVAILABILITY
EXAMPLE
The following instructs the oscilloscope to perform an auto-setup:
CMD$=“ASET”: CALL IBWRT(SCOPE%,CMD$)