P

A R T

T

W O

:

C O M M A N D S

204

ISSUED: February 2005

WM-RCM-E Rev D

MISCELLANEOUS

PASS_FAIL, PF

Command/Query

DESCRIPTION

The PASS_FAIL command sets up the pass/fail system.

COMMAND SYNTAX

PASS_FAIL

[,[,]]

:={ON,OFF}

:={AllTrue,AllFalse,AnyTrue,AnyFalse,
AllQ1ToQ4OrQ5ToQ8,AnyQ1ToQ4AndAnyQ5ToQ8}

:={1 to 1,000,000,000 sweeps}

QUERY SYNTAX

PASS_FAIL? or PF?

EXAMPLE (GPIB)

The following instruction sets the pass/fail system to all false, and

to stop after 20 acquisitions:

CMD$=”PF ON,ALLFALSE,20”: CALL
IBWRT(SCOPE%,CMD$)

RELATED COMMANDS

CRMS