DISPLAY

VERT_POSITION, VPOS

Command/Query

DESCRIPTION

The VERT_POSITION command adjusts the vertical position of
the specified trace on the screen. It does not affect the original

offset value obtained at acquisition time.

The VERT_POSITION? query returns the current vertical
position of the specified trace.

N

N

O

O

T

T

E

E

:

: The VPOS command and query can only be applied to

math function and memory traces. It does not apply to channel

inputs.

COMMAND SYNTAX

: Vert_POSITION

: = {F1,F2,F3,F4,F5,F6,F7,F8,TA,TB,TC,
TD,M1,M2,M3,M4

} TA through TD are included for backward

compatibility with software designed for earlier LeCroy instruments.

These four mnemonics will not be returned in response to queries.

: = -5900 to +5900 DIV

N

N

O

O

T

T

E

E

:

: The unit DIV is optional. The limits depend on the

current magnification factor, the number of grids on the

display, and the initial position of the trace.

QUERY SYNTAX

: Vert_POSition?

RESPONSE FORMAT

: Vert_POSITION

EXAMPLE

The following instruction shifts Trace A (TA) upwards by +3

divisions relative to the position at the time of acquisition:

CMD$=“F1:VPOS 3DIV”: CALL IBWRT(SCOPE%,CMD$)

RELATED COMMANDS

VERT_MAGNIFY