beautypg.com

Remote control commands and queries, Acquisition – Teledyne LeCroy X-STREAM OSCILLOSCOPES Remote Control User Manual

Page 259

background image

Remote Control Commands and Queries

WM-RCM-E Rev D

ISSUED: February 2005

253

ACQUISITION

WAIT

Command

DESCRIPTION

The WAIT command prevents your instrument from analyzing new
commands until the current acquisition has been completed. The

optional argument specifies the timeout (in seconds) after which the
scope will stop waiting for new acquisitions. If is not given, or if

= 0.0, the scope will wait indefinitely.

COMMAND SYNTAX

WAIT []

:=timeout in seconds (default is indefinite)

EXAMPLE (GPIB)

send: “TRMD SINGLE”

loop {send:”ARM;WAIT;C1:PAVA?MAX”

read response

process response

}

This example finds the maximum amplitudes of several signals
acquired one after another. ARM starts a new data acquisition. The
WAIT

command ensures that the maximum is evaluated for the

newly acquired waveform.

C1:PAVA?MAX

instructs the oscilloscope to evaluate the

maximum data value in the Channel 1 waveform.

RELATED COMMANDS

*TRG, TRIG_MODE, ARM