Remote Control Commands and Queries

WM-RCM-E Rev D

ISSUED: February 2005

251

ACQUISITION

VOLT_DIV, VDIV

Command/Query

DESCRIPTION

The VOLT_DIV command sets the vertical sensitivity in
Volts/div. The VAB bit (bit 2) in the STB register (see table on

page 225) is set if an out-of-range value is entered.

The probe attenuation factor is not taken into account for

adjusting vertical sensitivity.

The VOLT_DIV? query returns the vertical sensitivity of the
specified channel.

COMMAND SYNTAX

: Volt_DIV
: = {C1, C2, C3, C4}

: = See Operator’s Manual for specifications.

N

N

O

O

T

T

E

E

:

: The unit V is optional.

QUERY SYNTAX

: Volt_DIV?

RESPONSE FORMAT

: Volt_DIV

AVAILABILITY

: = {C3, C4} only available on four-channel
oscilloscopes.

EXAMPLE

The following instruction sets the vertical sensitivity of channel 1 to
50 mV/div:

CMD$=“C1:VDIV 50MV”: CALL IBWRT(SCOPE%,CMD$)