beautypg.com

Teledyne LeCroy X-STREAM OSCILLOSCOPES Remote Control User Manual

Page 104

background image

P

A R T

T

W O

:

C O M M A N D S

98

ISSUED: February 2005

WM-RCM-E Rev D

MISCELLANEOUS

*CAL?

Query

DESCRIPTION

The *CAL? query causes the oscilloscope to perform an internal
self-calibration and generates a response that indicates whether or

not your oscilloscope completed the calibration without error. This
internal calibration sequence is the same as that which occurs at

power-up. At the end of the calibration, after the response has
indicated how the calibration terminated, the oscilloscope returns to

the state it was in just prior to the calibration cycle. This includes the
AUTO_CALIBRATE status, which is not affected by the use of

*CAL?. That is, *CAL? may be used whether AUTO_CALIBRATE
has been set on or off.

QUERY SYNTAX

*CAL?

RESPONSE FORMAT

*CAL

: = 0 or other

0

= Calibration successful

EXAMPLE (GPIB)

The following instruction forces a self-calibration:

CMD$=“*CAL?”: CALL IBWRT(SCOPE%,CMD$):

CALL IBRD(SCOPE%,RD$): PRINT RD$

Response message (if no failure): *CAL 0

RELATED COMMANDS

AUTO_CALIBRATE