Well selectors, Cycles to analyze – Bio-Rad Firmware & Software Updates User Manual
Page 91
![background image](/manuals/600142/91/background.png)
Data Analysis Overview
78
Figure 54. Fluorophore analysis mode selected.
When Target is selected, data traces are displayed by target name as entered in the plate
setup (Figure 55).
Figure 55. Target analysis mode selected.
Cycles to Analyze
To restrict data analysis to a specified range of cycles, select Settings > Cycles to Analyze.
Select the starting cycle and the ending cycle using the arrow buttons or by typing in the
desired values and pressing Enter. Click the Restore Defaults button to return to the cycles
originally used for analysis.
NOTE: Removing cycles from the beginning of a run can have a significant impact
on baselining.
Well Selectors
Click the wells in the well selector to show or hide the data in the charts or spreadsheets
throughout the Data Analysis window:
•
To hide one well, highlight and click the individual well. To show that well, highlight
and click the well again