Chemical hazards, Explosive or flammability hazards, Electrical hazards – Bio-Rad Firmware & Software Updates User Manual

Page 7

background image

vi

No biohazardous substances are exhausted during normal operations of this instrument

S

URFACE

DECONTAMINATION

WARNING! To prevent electrical shock, always turn off and unplug the instrument prior to performing
decontamination procedures.

The following areas can be cleaned with any hospital-grade bactericide, virucide, or fungicide
disinfectant:

Outer lid and chassis

Inner reaction block surface and reaction block wells

Control panel and display

To prepare and apply the disinfectant, refer to the instructions provided by the product manufacturer.
Always rinse the reaction block and reaction block wells several times with water after applying a
disinfectant. Thoroughly dry the reaction block and reaction block wells after rinsing with water.

WARNING! Do not use abrasive or corrosive detergents or strong alkaline solutions. These agents can
scratch surfaces and damage the reaction block, resulting in loss of precise thermal control.

D

ISPOSAL

OF

B

IOHAZARDOUS

M

ATERIAL

The CFX96 Touch, CFX96 Touch Deep Well, CFX Connect, and CFX384 Touch systems contain no
potentially hazardous chemical materials. Dispose of the following potentially contaminated materials in
accordance with laboratory local, regional, and national regulations:

Clinical samples

Reagents

Used reaction vessels or other consumables that may be contaminated

Chemical Hazards

The CFX96 Touch, CFX96 Touch Deep Well, CFX Connect, and CFX384 Touch systems contain no
potentially hazardous chemical materials.

Explosive or Flammability Hazards

The CFX96 Touch, CFX96 Touch Deep Well, CFX Connect, and CFX384 Touch systems pose no
uncommon hazard related to flammability or explosion when used in a proper manner as specified by
Bio-Rad Laboratories.

Electrical Hazards

The CFX96 Touch, CFX96 Touch Deep Well, CFX Connect, and CFX384 Touch systems pose no
uncommon electrical hazard to operators if installed and operated properly without physical modification
and connected to a power source of proper specification.