Ow (figure 31), Plate editor menu bar – Bio-Rad Firmware & Software Updates User Manual
Page 59
Plates
46
Figure 31. Plate Editor window.
Plate Editor Menu Bar
The menu bar in the Plate Editor window provides the menu items shown in Table 15.
Table 15. Plate Editor menu bar.
Menu Item
Command
Function
File
Save
Save the plate files
Save As
Save the plate file with a new file name
Close
Close the Plate Editor
Settings
Plate Size
Select a plate size that reflects the number of wells in
the instrument block. Choose 384-well for the CFX384
Touch™ system or 96-well for the CFX96 Touch™,
CFX96 Touch Deep Well™, or CFX Connect™ system
NOTE: Plate Size must be the same as the block size in
the instrument on which the run will proceed.
Plate Type
Choose the type of wells in the plate that holds your
samples, including BR White and BR Clear. For
accurate data analysis, the plate type must be the same
as the plate well type used in the run
NOTE: You must calibrate new plate types (page 151).
Number Convention
Select or cancel the selection for Scientific Notation