Setting up the system – Bio-Rad Firmware & Software Updates User Manual
Page 17
![background image](/manuals/600142/17/background.png)
System Installation
4
Figure 3. Back panel of C1000 Touch thermal cycler.
WARNING! Avoid contact with the back panel of the C1000 Touch cycler during
operation.
Setting up the system
The CFX96 Touch, CFX96 Touch Deep Well, CFX Connect, or CFX384 Touch real-time PCR
detection system should be installed on a clean, dry, level surface with sufficient cool airflow to
run properly. The CFX96 Touch, CFX96 Touch Deep Well, or CFX384 Touch systems can be
run in two modes: stand-alone or software-controlled. The CFX Connect system can be run
only in software-controlled mode. If you are running the system under software-controlled
mode, make sure there is sufficient space for a computer during setup.
To insert the optical reaction module into the reaction module bay of the C1000 Touch or
CFX Connect chassis, follow these instructions:
1. Place the C1000 Touch or CFX Connect chassis in a suitable location with the locking
bar down. Remove any previously installed reaction modules.