INFICON SQC-310 Thin Film Deposition Controller User Manual
Page 90

5 - 10
IP
N 07
4-
55
0-
P1
B
SQC-310 Operating Manual
Command: . . . . . GB
Description: . . . . Gets/Sets the parameters of each of the 8 or 16 digital relays.
Parameters:
1
Name
4
Pulse Width (milliseconds)
2
Type (N.O. or N. C.)
5
Pin Number
3
Pulses
6
Relay In Use *
7
NameMode
Format:. . . . . . . . GB[Relay#]? [Parameter1] [Parameter2] …
Example: . . . . . . GB1? 2 5
Returns: !A2,0 5,1
NOTE: When changing the name (GB1), you MUST also send
the Name Mode command (GB7) as the next
parameter.
Command: . . . . . GC
Description: . . . . Gets the current state of each of the 8 or 16 digital relays and
inputs.
Parameters:
1
Current Relay State (off=0, on=1)
2
Current Input State (off=0, on=1)
Format:. . . . . . . . GC[RelayOrInput#][?] [Parameter1] [Parameter2] …
Example: . . . . . . GC1? 1 2
Returns: A1,0 2,1
Command: . . . . . GD
Description: . . . . Override the current relay state for one of the 8 or 16 digital
relays. Overriding the relay is only available while the instrument
is stopped. If a relay is accidentally left on, the relay is turned off
at the start of a new process.
Parameters:
1
Override Relay: Turn On
2
Override Relay: Turn Off
Format:. . . . . . . . GD[Relay#] [Parameter1]
Example: . . . . . . GD1 1
Returns: A
Command: . . . . . GE
Description: . . . . Allows any relay that is not controlled by a Logic Statement,
Sensor or Source, to be set remotely. Once a relay’s state has
been set remotely, that relay is locked and the instrument will not
allow it to be controlled via a Logic Statement Action, Sensor or
Source. Be sure to unlock the relay after it is no longer needed;
doing so will also turn the relay off.