7 get/set material parameters, 8 get/set input & relay parameters – INFICON SQC-310 Thin Film Deposition Controller User Manual
Page 89

5 - 9
IP
N 07
4-
55
0-
P1
B
SQC-310 Operating Manual
5.3.7 Get/Set Material Parameters
Command: . . . . . F
Description: . . . . Gets/Sets the parameters of the 100 stored materials.
Parameters:
1.
Material Name
3.
Zfactor
2.
Density
Example: . . . . . . To Get Material 1 name, send: F1? 1
The return string would be of the form: Aluminum
To Set the parameter you would send: F1 1,AnyName
NOTE: You can only get/set one material parameter at a time.
5.3.8 Get/Set Input & Relay Parameters
Command: . . . . . G
Description: . . . . Gets/Sets the logical function of each of the 16 digital inputs.
Parameters:
1
Start Process
32
Soak Hold
2
Abort Process
33
Zero Thickness
3
Start Layer
34
Zero Time
4
Stop Layer
35
Out1 Pocket Ready
5
Start Next Layer
36
Out2 Pocket Ready
6
Force Final Thick
37
Out3 Pocket Ready
7-31
Start Process 1-25
38
Out4 Pocket Ready
39
Not Used
Example: . . . . . . To Get Input 1 and Input 2 functions, send: G? 1 2
The return string for Start Layer and Stop Layer would be:
1,3 2,4
To Set the same parameters you would send: G1 1,3 2,4
Command: . . . . . GA
Description: . . . . Gets/Sets the parameters of each of the 8 or 16 digital inputs.
Parameters:
1
Name
3
Pin Number
2
Active Level
4
Input In Use *
5
NameMode
Format: . . . . . . . . GA[Input#]? [Parameter1] [Parameter2] …
Example: . . . . . . GA1? 2 3
Returns: !A2,0 3,1
NOTE: When changing the name (GA1), you MUST also send
the Name Mode command (GA5) as the next
parameter.