SQC-310 Operating Manual

A4:

1

Shutter Delay

5

Rate Sampling (0,1,2)

2

Capture

6

Sample Accuracy

3

Control Error (0,1,2)

7

Sample Hold

4

Control Percent

8

Sample Time

A5:

1

Snsr 1 Crystal Fail Mode

9

Snsr 3 Crystal Fail Mode

2

Snsr 1 Crystal Position

10

Snsr 3 Crystal Position

3

Snsr 1 Backup Sensor

11

Snsr 3 Backup Sensor

4

Snsr 1 BackupXtalPosition

12

Snsr 3 BackupXtalPosition

5

Snsr 2 Crystal Fail Mode

13

Snsr 4 Crystal Fail Mode

6

Snsr 2 Crystal Position

14

Snsr 4 Crystal Position

7

Snsr 2 Backup Sensor

15

Snsr 4 Backup Sensor

8

Snsr 2 BackupXtalPosition

16

Snsr 4 BackupXtalPosition

Example: . . . . . . To Get the PID parameters of Film 1, send: A2 1? 1 2 3

The return string would be of the form: 1,50 2,5 3,0

To Set the PID parameters of Film 1, send: A2 1 1,50 2,5 3,0

NOTE: Numeric film parameter values sent and received by

the SQC-310 are integers. To convert between the
integer value and the parameter’s actual value, you
must multiply or divide by the number of decimal digits
in the displayed parameter.

In the examples above, the P term is displayed as an integer
value, so no conversion is required. However, the I Term is
displayed with one decimal digit (i.e., 0.5). A Get response of 5
actually represents a parameter value of .5. Similarly, a Set
value of 5 for the I Term sets its value in the SQC-310 to .5.