Auxiliary functions

26.2.15.3. Resetting the settings under

CAL.USER to the factory settings

Key

Action

Description

MENU

Press

for approx. 3 s

Switching from process level

setting level.

/

Select

CAL.USER

(To do this, the auxiliary function must be incorporated into the main
menu).

ENTER

The submenu options are displayed.

/

Select

copy FACT->USER

RUN

Hold down

as long as countdown (5 ...) is
running

The settings of

CAL.USER are reset to the factory settings.

EXIT

Press

Acknowledgment and simultaneous return to the main menu (MAIN).

EXIT

Press

Switching from setting level

process level.

Table 66:

copy FACT->USER; resetting the settings under CAL.USER to the factory settings

The factory calibration is re-activated by deactivating

CAL.USER, by removing the auxiliary function from

the main menu (MAIN).