beautypg.com

Foxy-t1000 specifications – Ocean Optics OOISensors User Manual

Page 94

background image

6: Hardware Data Sheets and Instructions

FOXY-T1000 Specifications

Specification Value

Fiber core

1000

µm silica

Fiber cladding

Silica

Fiber jacketing

Stainless steel

Outer diameter

1/4"

Length 7"

Amount of pressure that can be
applied

3000 psi

Probe temperature range

-80

° C to 80° C (reflects range of ruthenium complex in the sol-

gel matrix)

Connector SMA

905

Response time (without overcoat)

<1 second

Compensation Temperature

compensation

only

Probe reconditioning

Yes (FOXY-RECOV @ $100)

Re-calibration

Required when probe is replaced or cleaned by autoclave

Probe lifetime

1 year

FOXY-T1000-RTD Stainless-steel Fiber Optic
Probe with Light Shield

The FOXY-T1000-RTD is a 1000-

µm silica core, silica-clad fiber encased in a stainless-steel ferrule with

a screw-on light shield. The light shield prevents ambient light from entering the probe. The 7" long probe
has an outer diameter of 1/4" and is extremely rugged. The probe can withstand pressure up to 3000 psi
and temperature up to 80

° C. It is designed for use with the 600-µm bifurcated optical fiber assembly.

See

Oxygen Sensing Probes – General Information

at the beginning of this chapter for information on

cleaning and sterilizing oxygen probes.

84

FOXY-AL300-000-02-0207