Foxy-t1000 specifications – Ocean Optics OOISensors User Manual
Page 94
6: Hardware Data Sheets and Instructions
FOXY-T1000 Specifications
Specification Value
Fiber core
1000
µm silica
Fiber cladding
Silica
Fiber jacketing
Stainless steel
Outer diameter
1/4"
Length 7"
Amount of pressure that can be
applied
3000 psi
Probe temperature range
-80
° C to 80° C (reflects range of ruthenium complex in the sol-
gel matrix)
Connector SMA
905
Response time (without overcoat)
<1 second
Compensation Temperature
compensation
only
Probe reconditioning
Yes (FOXY-RECOV @ $100)
Re-calibration
Required when probe is replaced or cleaned by autoclave
Probe lifetime
1 year
FOXY-T1000-RTD Stainless-steel Fiber Optic
Probe with Light Shield
The FOXY-T1000-RTD is a 1000-
µm silica core, silica-clad fiber encased in a stainless-steel ferrule with
a screw-on light shield. The light shield prevents ambient light from entering the probe. The 7" long probe
has an outer diameter of 1/4" and is extremely rugged. The probe can withstand pressure up to 3000 psi
and temperature up to 80
° C. It is designed for use with the 600-µm bifurcated optical fiber assembly.
See
Oxygen Sensing Probes – General Information
at the beginning of this chapter for information on
cleaning and sterilizing oxygen probes.
84
FOXY-AL300-000-02-0207