6: Hardware Data Sheets and Instructions

FOSPOR Specifications

Parameter

Dissolved Oxygen in Liquid

Oxygen Gas (at 1 atmosphere)

Sensor mechanism

Phase shift due to change in partial pressure of O

2

Dynamic range

0–40.7 ppm, 0–760 mmHg partial
pressure

0–100 mole%, 0–760 mmHg partial
pressure

Response time

Standard: ~1 sec
20–50 seconds with silicone
overcoating

Standard: ~1 sec
20–30 seconds with silicone
overcoating

Resolution

0.002 ppm at room temperature

0.005% at room temperature

Stability

Drift < 0.00008 ppm O

2

per hour at

low DO

Drift ~ 0.006 ppm O

2

per hour in

air-saturated water

Drift ~ 0.0002% O

2

drift per hour at

low O

2

Drift ~ 0.015% O

2

per hour air

Minimum detectable limit

0.002 ppm at room temperature

0.005 mole% (0.04 mm Hg)

Accuracy (0-20% O

2

, 0-50ºC)

5% of reading (using 2

nd

order polynomial fit to multipoint calibration)

Temperature

Temperature sensor and software-supported compensation

Temperature range

0–60ºC

Probe reconditioning

HIOXY-RECOV: $150

HIOXY-RECOV-N (for HIOXY needle probes: $175

Calibration

Standard 2-point calibration (linear fit)

Multipoint calibration (2

nd

order polynomial fit) for improved accuracy and

broad dynamic range applications

Single-temperature, single point calibration using factory-calibrated oxygen
probes

Temperature recalibration

Only when probe is replaced/refurbished

Storage conditions

No specific requirements

Cleaning

10% hypochlorite detergent

Sterilization

Gamma radiation or sodium hypochloride (bleach or Clorox)

FOXY-AL300-000-02-0207