Fospor specifications – Ocean Optics OOISensors User Manual
Page 115
6: Hardware Data Sheets and Instructions
FOSPOR Specifications
Parameter
Dissolved Oxygen in Liquid
Oxygen Gas (at 1 atmosphere)
Sensor mechanism
Phase shift due to change in partial pressure of O
2
Dynamic range
0–40.7 ppm, 0–760 mmHg partial
pressure
0–100 mole%, 0–760 mmHg partial
pressure
Response time
Standard: ~1 sec
20–50 seconds with silicone
overcoating
Standard: ~1 sec
20–30 seconds with silicone
overcoating
Resolution
0.002 ppm at room temperature
0.005% at room temperature
Stability
Drift < 0.00008 ppm O
2
per hour at
low DO
Drift ~ 0.006 ppm O
2
per hour in
air-saturated water
Drift ~ 0.0002% O
2
drift per hour at
low O
2
Drift ~ 0.015% O
2
per hour air
Minimum detectable limit
0.002 ppm at room temperature
0.005 mole% (0.04 mm Hg)
Accuracy (0-20% O
2
, 0-50ºC)
5% of reading (using 2
nd
order polynomial fit to multipoint calibration)
Temperature
Temperature sensor and software-supported compensation
Temperature range
0–60ºC
Probe reconditioning
HIOXY-RECOV: $150
HIOXY-RECOV-N (for HIOXY needle probes: $175
Calibration
Standard 2-point calibration (linear fit)
Multipoint calibration (2
nd
order polynomial fit) for improved accuracy and
broad dynamic range applications
Single-temperature, single point calibration using factory-calibrated oxygen
probes
Temperature recalibration
Only when probe is replaced/refurbished
Storage conditions
No specific requirements
Cleaning
10% hypochlorite detergent
Sterilization
Gamma radiation or sodium hypochloride (bleach or Clorox)
FOXY-AL300-000-02-0207