Foxy/hoxy/fospor general specifications – Ocean Optics OOISensors User Manual
Page 114

6: Hardware Data Sheets and Instructions
FOXY/HOXY/FOSPOR General Specifications
Specification FOXY
HIOXY
FOSPOR
Principle Photoluminescence-
quenching using Ru
compound. Measures O
2
partial pressure.
Photoluminescence-
quenching using Ru
compound. Measures O
2
partial pressure.
Photoluminescence-
quenching using Pt
compound. Measures O
2
partial pressure.
Applications Benign
environment
Aqueous liquids and vapors
Hydrocarbon-based
liquids and vapors
Fuels
Alcoholic beverages
Vegetable oil
Low O
2
(0–5%)
Vacuum systems
packaging
Time constant
(µsec) dynamic
range
Air: 2.50 N
2
: 5.00
Low: 1.00 High: 7.00
Air: 0.50 N
2
: 2.50
Low: 0.30 High: 6.00
Air: 10.00 N
2
: 50.00
Low: 3.00 High: 70.00
Temperature
range
-50ºC to +80ºC
-50ºC to +80ºC
0ºC to +60ºC
Response time
~1 sec – liquid/gas no
overcoating
15–20 sec in gas-
overcoated
30–45 sec in liquid-
overcoated
~1 sec – liquid and gas
~1 sec – liquid/gas no
overcoating
15–20 sec in gas-
overcoated
30–45 sec in liquid-
overcoated
Overcoating
With or without overcoating
No overcoating
With or without overcoating
104
FOXY-AL300-000-02-0207