6: Hardware Data Sheets and Instructions
HIOXY/MFPF Specifications
Specification
Dissolved Oxygen in Liquid
Oxygen Gas (at 1 atmosphere)
Sensor mechanism
Phase shift due to change in partial pressure of O
2
Dynamic range
0–40.7 ppm
0–760 mm Hg partial pressure
0–100% (mole percent)
0–760 mm Hg partial pressure
Response time (standard)
~ 1 second
Response time (w/overcoat)
Not applicable
Temperature
Temperature sensor and software-supported compensation
Temperature range
-50º C to 80º C
Chemical compatibility of
coating
Compatible with variety of hydrocarbon-based fluids and vapors such as
solvents, fuels and oils
Tested with jet fuel, diesel, gasoline, hydraulic fluids, ethanol, and wine
Resolution
0.02 ppm at room temperature 0.05% (0.4 mm Hg) at room temperature
Calibration
Standard 2-point calibration (linear fit)
Multipoint calibration (2
nd
order polynomial fit) for improved accuracy and
broad dynamic range applications
Single-temperature, single-point calibrations using factory-calibrated
FOXY probes
Accuracy: (0–20% O
2
, 0–50ºC)
5% of reading (using 2
nd
order polynomial fit to multipoint calibration)
Lowest detectable limit
0.02 ppm at room temperature
0.05% (0.4 mm Hg)
Stability
Drift<0.002 ppm per hour
Drift ~0.01% O
2
per hour
Probe reconditioning
HIOXY-RECOV $150
HIOXY-RECOV-N (for HIOXY needle probes) $175
Temperature recalibration
Only when probe is replaced/refurbished
Storage conditions
No specific requirements
Cleaning
10% hypochlorite detergent
Sterilization
Gamma radiation or sodium hypochlorite (bleach or Clorox)
Probe lifetime
Refurbishing recommended once per year
FOXY-AL300-000-02-0207