6: Hardware Data Sheets and Instructions

HIOXY/MFPF Specifications

Specification

Dissolved Oxygen in Liquid

Oxygen Gas (at 1 atmosphere)

Sensor mechanism

Phase shift due to change in partial pressure of O

2

Dynamic range

0–40.7 ppm
0–760 mm Hg partial pressure

0–100% (mole percent)
0–760 mm Hg partial pressure

Response time (standard)

~ 1 second

Response time (w/overcoat)

Not applicable

Temperature

Temperature sensor and software-supported compensation

Temperature range

-50º C to 80º C

Chemical compatibility of
coating

Compatible with variety of hydrocarbon-based fluids and vapors such as
solvents, fuels and oils

Tested with jet fuel, diesel, gasoline, hydraulic fluids, ethanol, and wine

Resolution

0.02 ppm at room temperature 0.05% (0.4 mm Hg) at room temperature

Calibration

Standard 2-point calibration (linear fit)

Multipoint calibration (2

nd

order polynomial fit) for improved accuracy and

broad dynamic range applications

Single-temperature, single-point calibrations using factory-calibrated
FOXY probes

Accuracy: (0–20% O

2

, 0–50ºC)

5% of reading (using 2

nd

order polynomial fit to multipoint calibration)

Lowest detectable limit

0.02 ppm at room temperature

0.05% (0.4 mm Hg)

Stability

Drift<0.002 ppm per hour

Drift ~0.01% O

2

per hour

Probe reconditioning

HIOXY-RECOV $150

HIOXY-RECOV-N (for HIOXY needle probes) $175

Temperature recalibration

Only when probe is replaced/refurbished

Storage conditions

No specific requirements

Cleaning

10% hypochlorite detergent

Sterilization

Gamma radiation or sodium hypochlorite (bleach or Clorox)

Probe lifetime

Refurbishing recommended once per year

FOXY-AL300-000-02-0207