4: pH Sensors

Note

Be careful not to over-tighten the tip. Over-tightening the tip can break the mirror in the
transmission probe or break the film in the reflection probe.

6. Place the probe in the first pH sample, ensuring that there are no bubbles in the light path of the

probe (only applicable to transflectance probes).

pH Sensors Analysis Wavelengths

Indicator Film

Valid pH
Range

Isobestic
Wavelength
(nm)

HA Peak
Wavelength
(nm)

A Peak
Wavelength
(nm)

Baseline
Wavelength (nm)

Phenol Red

6.5 – 8.5

470

435

575

750

Cresol Red

8.0 – 10.0 490

424

587

750

m-Cresol Purple

8.5 – 10.5 487

435

593

800

Thymol Blue

9.0 – 12.0 495

440

605

800

Brilliant Yellow

7.0 – 9.0

465

412

523

700