beautypg.com

Ocean Optics OOISensors User Manual

Page 59

background image

3: Oxygen Sensors

6. Get the proper temperature reading. Either:

• Select Spectrometer | Sample Temperature from the menu bar in the Single Temperature

Calibration screen, or

• Enter the temperature manually.

7. Leave the oxygen probe in the standard for at least 5 minutes. This guarantees equilibrium.

8. Place the cursor in the Intensity box.

9. Click the green Scan Standard button or select Spectrometer | Scan Standard from the menu

bar.

Note

Enabling the optional Continuous function, located to the right of the Scan Standard button,
will allow you to receive continuous intensity values of a standard. To use this function, check the
Continuous box.

• Once you click the green Scan Standard button, a red Scanning button appears. Watch the

values in the Intensity column (If this value is < 1.0, then the Enable reference correction
option is checked on the Configure Software screen.). When there appears to be no changes
in this value, select the red Scanning button to accept the intensity value.

FOXY-AL300-000-02-0207

49