Usb-ls-450 temperature measurement specifications, Usb-ls-450 power requirements, Bifurcated optical fiber assemblies – Ocean Optics OOISensors User Manual
Page 112
6: Hardware Data Sheets and Instructions
USB-LS-450 Temperature Measurement Specifications
Specification Value
Precision
Better than 0.1° C
Accuracy
Better than 0.5° C
Maximum data rate
15 samples/second
USB-LS-450 Power Requirements
Specification Value
Input voltage range
3.0-8.0 volts
Quiescent range
~20 mA
Current with light on
~20 mA
Bifurcated Optical Fiber Assemblies
Ocean Optics’ bifurcated fiber optic assemblies easily connect to oxygen sensors, light sources, and
miniature fiber optic spectrometers via SMA terminations and an additional splice bushing. These single-
strand, multi-mode optical fibers are silica-core and silica-clad. Standard assemblies are 2 meters in
length, although custom fibers can be greater than 2 meters. The ends of the fibers are cleaved, epoxied
into the connectors, and polished.
Bifurcated assemblies are shaped like a "Y" with a stainless steel breakout located midway from the ends
of the fibers. The common end (the tail of the "Y") of a bifurcated assembly has two fibers side by side.
The common end of a bifurcated fiber can be coupled to a larger diameter probe with a splice bushing.
Bifurcated fibers used in Fiber Optic Sensors System include the following:
Bifurcated Fiber
Description
BIF200
2 separate 200-
µm optical fibers in a "Y" configuration. For use with the
FOXY-AL300 and FOXY-24G probes
BIF400
2 separate 400-
µm optical fibers in a "Y" configuration. For use with the
FOXY-PI600 probe
BIF600
2 separate 600-
µm optical fibers in a "Y" configuration. For use with the
FOXY-R, FOXY-OR125, FOXY-OR125G and the FOXY-T1000 probes
102
FOXY-AL300-000-02-0207