Calibrating standards – Ocean Optics OOISensors User Manual
Page 55
3: Oxygen Sensors
9. Select the algorithm type from the Curve Fitting drop-down menu. This algorithm will be used
to calibrate your sensor system. See the
Calibration Requirements for Oxygen Sensors
more information on algorithms
You have configured the system for calibration. Proceed to the next section to calibrate your standards.
Calibrating Standards
►
Procedure
1. Enter information for the first standard in the Calibration Table. The first standard should have
0% oxygen concentration, such as in a nitrogen flow or in a solution of sodium hydrosulfite.
Enter 0 in the Concentration column of the Calibration Table.
2. Leave the oxygen probe in the standard for at least 5 minutes. This guarantees equilibrium.
3. Move your cursor to the Intensity column.
4. Click the green Scan Standard button or select Spectrometer | Scan Standard from the menu.
Note
Enabling the optional Continuous function, located to the right of the Scan Standard button,
will allow you to receive continuous intensity values of a standard. To use this function, check the
Continuous box.
FOXY-AL300-000-02-0207