beautypg.com

Calibrating standards using factory calibration – Ocean Optics OOISensors User Manual

Page 58

background image

3: Oxygen Sensors

3. Adjust the integration time so that the fluorescence peak does not exceed 3500 counts. Set the

integration time to powers of two (2, 4, 8, 16, 32, 64 etc.) to ensure a constant number of LED
pulses during the integration time. (The intensity of the LED peak [~475 nm] does not affect your
measurements.)

4. Select

Calibrate | Oxygen, Multiple Temperature from the menu bar. The Multiple

Temperature Calibration screen appears.

5. Select

File | Open Calibration Table from the menu bar. The name of the file corresponds to the

serial number of the probe that you are calibrating. Once this file is opened, the Calibration Table
should be populated with oxygen concentration amounts and temperatures.

Note

The serial number is etched on the SMA connector, located at the top of the probe.

6. Select the green Curve Fit button. Graphs displaying the curves appear in the bottom of the

dialog box. Verify that the graphs are satisfactory, and then proceed to Step 7.

7. Click the green Update Channel Calibration button to save information from this calibration

procedure.

8. Select

File | Close from the menu bar to return to the main display window.

You have now configured the system for calibration using the factory calibration file. Proceed to the next
section to calibrate your standards.

Note

Second order is strongly recommended for all temperature compensation.

Calibrating Standards Using Factory Calibration

Procedure

1. Select

Calibrate | Oxygen, Single Temperature from the OOISensors menu bar.

2. Click on the Calibration Type drop-down menu and select Single Point.

3. Click on the Curve Fitting drop-down menu and select Second Order Polynomial.

4. Enter the known oxygen concentration of your standard under Concentration.

5. Change the switch in the Temp Compensation section of the Single Temperature Calibration

screen to Yes.

48

FOXY-AL300-000-02-0207