6: Hardware Data Sheets and Instructions

FOXY/MFPF Specifications

Specification

Dissolved Oxygen in H

2

O

Oxygen Gas (at 1 atmosphere)

Sensor mechanism

Phase shift due to change in partial pressure of O

2

Dynamic range

0–40.7 ppm
0–760 mm Hg partial pressure

0–100% (mole percent)
0–760 mm Hg partial pressure

Response time (standard)

~ 1 second

Response time (w/overcoat)

30–50 seconds

20–30 seconds

Temperature

Temperature sensor and software-supported compensation

Temperature range

-50º C to 80º C

Chemical compatibility of coating Unaffected by pH change or salinity

Safe with 50% methanol, acids, sodium sulfite, moisture, CO

2

, and CO

Not recommended with strong bases (pH>10), styrene, ethanol, liquid
acetone, acetonirile, HF and BTEX solvents

Resolution

0.02 ppm at room temperature

0.05% (0.04 mmHg) at room
temperature

Calibration

Standard 2-point calibration (linear fit)
Multipoint calibration (2

nd

order polynomial fit) for improved accuracy

and broad dynamic range applications
Single-temperature, single-point calibrations using factory-calibrated
oxygen probes

Accuracy: (0–20% O

2

, 0–50ºC)

5% of reading (using 2

nd

order polynomial fit to multipoint calibration)

Lowest detectable limit

0.02 ppm

0.05% (0.04 mm Hg)

Stability

Drift<0.01 ppm per hour

Drift ~0.03% O

2

per hour

Probe reconditioning

FOXY-RECOV $100
FOXY-RECOV-N (for FOXY needle probes) $125

Temperature recalibration

Only when probe is replaced/refurbished

Storage conditions

No specific requirements

Cleaning

10% hypochlorite detergent for overcoated probes

Sterilization

Gamma radiation or sodium hypochlorite (bleach or Clorox)

Probe lifetime

Refurbishing recommended once per year

106

FOXY-AL300-000-02-0207