beautypg.com

Calibrate menu functions, Configuring a spectrometer in ooisensors – Ocean Optics OOISensors User Manual

Page 44

background image

2: OOISensors Software

Calibrate Menu Functions

The following section provides information on options available from the Spectrometer menu.

Option Name

Description

Oxygen, Single Temperature

Allows for oxygen probe calibration at a single temperature. Used
most frequently.

Oxygen, Multiple Temperature

Allows for oxygen probe calibration at multiple temperatures.

pH

Allows for pH probe calibration.

Pressure

Allows for pressure probe calibration.

Configuring a Spectrometer in OOISensors

Once you install OOISensors, you will need to configure the spectrometer in OOISensors using the
Wavelength Calibration data included with the spectrometer shipment.

Procedure

Perform the steps below to configure the spectrometer in OOISensors:

1. Select

Configure | Spectrometer from the menu bar. The Configure Spectrometer screen

appears.

2. Click on the Sensors tab, if necessary.

3. Ensure that the values in the Wavelength Calibration Coefficients section of this screen are the

same as the values on the Wavelength Calibration Data you received with the spectrometer.

4. Repeat Step 3 for each channel in the spectrometer. In the left pane of the screen, select the slave

channel number, then verify and/or enter the values in the Wavelength Calibration Coefficients
section of the screen.

34

FOXY-AL300-000-02-0207