About factory calibration, Calibrating with factory calibration – Ocean Optics OOISensors User Manual

Page 57

background image

3: Oxygen Sensors

Calibrating Oxygen Sensors with Temperature
Compensation

Perform the steps below to calibrate your oxygen sensors using temperature compensation.

About Factory Calibration

If you want to compensate for temperature changes, you can purchase a Factory Calibration to be
performed at Ocean Optics prior to shipping the Fiber Optic Sensors System. Consult an Ocean Optics
catalog, contact an Ocean Optics Applications Scientist, or visit our website at

www.oceanoptics.com

for a price for this service.

Note

Factory Calibration can only be performed if you specify the oxygen concentration range and
temperature in which you will be working when you purchase the Fiber Optic Sensors System.

Calibrating With Factory Calibration

Perform the steps below to calibrate the oxygen sensors for temperature compensation using the factory
calibration.

Note

If you are not using (or do not have) a factory calibration file, proceed to

Calibrating Without

Factory Calibration

later in this chapter.

Preparing for Calibration Using the Factory Calibration

If you ordered the Factory Calibration, you are provided with an additional file that includes data for the
Calibration Table. You will need this file in Step 5.

Procedure

1. Set the data acquisition parameters for the calibration procedure (integration time, averaging, and

boxcar smoothing).

2. Set the integration time for the entire calibration procedure when the probe is measuring the

standard with zero concentration at the lowest sample temperature you will be using. The
fluorescence peak (~600 nm) will be at its maximum at zero concentration.

FOXY-AL300-000-02-0207

47