2: OOISensors Software

Option Name

Description

External Software Trigger: This mode allows data acquisition to be
synchronized with external events. In this level-triggered mode, the
spectrometer is free running, only while level is high, just as in normal
mode. However, upon each trigger, the data collected up to the trigger
event is transferred to the software. For more information, consult the
External Triggering document on the Ocean Optics web site at:

http://www.oceanoptics.com/technical/externaltriggering.pdf

Display Tab

Click on the Display Tab in the Configure Spectrometer screen to access Display controls. This screen
allows you to configure display parameters for spectrometers in OOISensors. This tab does not exist for
systems with MFPF units.

The Display Tab contains the following options:

Option Name

Description

Show software controllable
LED status

Displays the status of the LS-450’s blue LED.

Graph and Chart Display
Mode

Specifies the information that appears in the display window.

Spectral Graph Only: The Spectral Graph appears in the display
window.