beautypg.com

Operation, Foxy-tk1 specifications – Ocean Optics OOISensors User Manual

Page 102

background image

6: Hardware Data Sheets and Instructions

Operation

Procedure

1. Connect the FOXY-TK1 to the 10-foot wire (model number FOXY-T-WIRE) that comes with the

thermocouple.

2. Connect the FOXY-TK1 and the wire to the FOXY-T-MOD-K RS-232 module. (Rack-mounted

systems have the modules installed into the rack box.)

3. Connect the FOXY-T-MOD-K RS-232 module to your PC. You must have an available serial

port on your PC. Note the serial port number (also called COM Port) on the PC to which you are
interfacing.

4. Plug the +12VDC wall transformer into an outlet and connect it to the FOXY-TK1.

5. Open OOISensors and select Configure | Spectrometer from the menu. Select the Sensors tab.

Next to Temperature Measurement, select Omega D5xx1 RS232 for automatic temperature
monitoring. Finally, enable the Compensate function.

6. Select the COM Port number (next to Serial Port) on your PC to which the thermocouple is

connected.

7. Locate the port on the module that is connected to the thermistor in use.

Note

The RS-232 module can support up to four thermocouples, and each port on the module is
labeled.

8. Next

to

D5xx1, select between 0 and 3. (If you only have one thermocouple, select 0.)

9. Enable

the

Chart function to view a chart of the temperature values.

FOXY-TK1 Specifications

Specification Value

Length 6"

Outer diameter

1/8"

Temperature range

-150

° C to 220° C

Accuracy

2.2

° C or 0.75% of reading with precision of 1° C

92