Use warnings, Oxygen probe specifications, Sterilization methods tested on oxygen probes – Ocean Optics OOISensors User Manual
Page 85
6: Hardware Data Sheets and Instructions
Use Warnings
• Handle with care. Dropping the probe may cause permanent damage.
• Gently remove the plastic cover from the SMA connector before use.
• Pulling the SMA connector away from the probe when removing the plastic cover could damage the
probe.
Oxygen Probe Specifications
The following specifications pertain to all oxygen probes:
Specification
Dissolved Oxygen in H
2
0
Oxygen Gas (at 1 atmosphere)
Dynamic range
0-40.7 ppm
0-760 mm Hg partial pressure
0-100% (mole percent)
0-760 mm Hg partial pressure
Response time (without
overcoat)
< 1 second
< 1 second
Response time (with
overcoat)
30-50 seconds
20-30 seconds
Resolution
0.003 ppm at 0 ppm O2
0.02 ppm at 8.5 ppm O2
0.2 ppm at 40 ppm O2
0.01% at 0% O2
0.04% at 20.9% O2
0.40% at 100% O2
Stability
Drift <0.02 ppm per day
Drift <0.05% per day
Minimum Detection Limit
0.01 ppm
0.025%
Sterilization Methods Tested on Oxygen Probes
The following sterilization methods have been tested for use with oxygen probes. The results of each
sterilization method are listed to in the right column of the table:
Method Results
Autoclaving (steam sterilization 30 minutes or
more at 121 °C)
Each cycle decreases signal by more than 50%.
Recalibration is necessary; lifetime of probe is 6- 8
cycles.
Gamma Radiation
No effect.
FOXY-AL300-000-02-0207