Use warnings, Oxygen probe specifications, Sterilization methods tested on oxygen probes – Ocean Optics OOISensors User Manual

Page 85

background image

6: Hardware Data Sheets and Instructions

Use Warnings

• Handle with care. Dropping the probe may cause permanent damage.
• Gently remove the plastic cover from the SMA connector before use.
• Pulling the SMA connector away from the probe when removing the plastic cover could damage the

probe.

Oxygen Probe Specifications

The following specifications pertain to all oxygen probes:

Specification

Dissolved Oxygen in H

2

0

Oxygen Gas (at 1 atmosphere)

Dynamic range

0-40.7 ppm
0-760 mm Hg partial pressure

0-100% (mole percent)
0-760 mm Hg partial pressure

Response time (without
overcoat)

< 1 second

< 1 second

Response time (with
overcoat)

30-50 seconds

20-30 seconds

Resolution

0.003 ppm at 0 ppm O2

0.02 ppm at 8.5 ppm O2

0.2 ppm at 40 ppm O2

0.01% at 0% O2

0.04% at 20.9% O2

0.40% at 100% O2

Stability

Drift <0.02 ppm per day

Drift <0.05% per day

Minimum Detection Limit

0.01 ppm

0.025%

Sterilization Methods Tested on Oxygen Probes

The following sterilization methods have been tested for use with oxygen probes. The results of each
sterilization method are listed to in the right column of the table:

Method Results

Autoclaving (steam sterilization 30 minutes or
more at 121 °C)

Each cycle decreases signal by more than 50%.
Recalibration is necessary; lifetime of probe is 6- 8
cycles.

Gamma Radiation

No effect.

FOXY-AL300-000-02-0207