Difference check, Setting method, Chapter 4 function settings – KEYENCE LS-7000 User Manual

Page 68: Select the area to be set and press the, Press the, Key three times, Press the or key and set the averaging number, Press the or key and set the difference value

background image

4-16

Chapter 4 Function Settings

4

Difference Check

If there is a radical change between measured values, the change will be detected and
output.

Setting Range
• Difference average: 1 to 16 times
• Difference value: 1 to 9999

μm

Note:

The results of difference checks will be output externally, provided that DIFF
function output settings are made properly in advance. Refer to Function
Output Selection on page 4-19 for details.
The result output of a difference check is a 26.6-ms-long single output signal,
which is not displayed on the control panel.

Setting Method

1

Select the area to be set and press
the

ENT

key.

2

Press the

FUNC

key three times.

3

Press the

or

key and set the averaging number.

The setting
part flashes.

Press these
keys to set
the value.

Averaging times

PROG No.

PRINT

AREA

DISP

ESC

PROG

FUNC

ENT

H

ZE

FOCUS

2

1

A R E A

POSITION

G

L

4

Press the

ENT

key.

The input value is entered and the difference value display flashes.

5

Press the

or

key and set the difference value.

PROG No.

PRINT

AREA

DISP

ESC

PROG

FUNC

ENT

H

ZE

FOCUS

2

G

L

The setting
part flashes.

Difference
value

Press these keys
to set the value.

HI

GO

LO