Mounting and Connection
Mount the measuring head and the controller, and connect them with the cables.
■ Mounting the Measuring Head
The measuring head can be mounted with or without the TR base employed, according
to the type of targets and operating environment.
The following section provides information on how to mount the measuring head with or
without the TR base employed, along with mounting conditions required.
■ Mounting the Controller
Mount the controller to panels, such as control or operation panels, or to the LS-S11 AC
Power Supply Stand. The LS-S11 Power Supply Stand is sold separately.
■ Connection
After the measuring head and the controller are mounted, connect them with the ca-
bles.
Mounting the Measuring Head
■ Mounting Restrictions
The transmitter and receiver must abide by the following restrictions when the measur-
ing head is mounted.
Vertical Position Mismatching
Angle Mismatching
±1 mm max.
Transmitter
Receiver
Transmitter
Receiver
±1 mm max.
Transmitter
Receiver
±
0.25
° max
Transmitter
Receiver
±
0.25
° max
Transmitter
Receiver
Transmitter
Receiver
±
0.25
° max
±
0.25
° max
The values are the same
for LS-7010, LS-7030, and
LS-7070.