Connection – KEYENCE LS-7000 User Manual
Page 31
1-17
1
Chapter 1 Getting Started
CAUTION
The back of the AC power supply
stand has legs. Erect the legs and
use the AC power supply stand if
required.
Tips
Connection
Connect the measuring head, controller, and power supply with the cables.
Do not supply power to the controller while connecting the
cables. If the AC power supply stand is used, turn off the AC
power supply stand while connecting the cables. The measur-
ing head, controller, or peripheral devices may be damaged if
power is supplied while connecting the cables.
Make the following three connections.
• Transmitter and Receiver:
Connect the cables from the transmitter and receiver
respectively.
• Measuring head and controller: Connect the cable between the controller and measur-
ing head.
• Power supply cable:
Supply 24 VDC to the controller directly or through the
AC power supply stand.
Power supply cable to the AC
power supply stand
Cable between controller
and measuring head
(LS-C3A/LS-C10A/LS-C30A)
Up to two cables are connectable,
provided that the total length of
the cables does not exceed 40 m.
Cable between controller
and measuring head
(LS-C3A/LS-C10A/LS-C30A)
Up to two cables are connectable,
provided that the total length of
the cables does not exceed 40 m.
KZ-U3 24-VDC Power Supply
(Optional)
Measuring head
(LS-7010/LS-7030/LS-7070)
LS-7001
Measuring head
(LS-7010/LS-7030/LS-7070)
Cable between transmitter and receiver
(OP-42182/OP-42183) (Optional)
Up to two cables are connectable,
provided that the total length of
the cables does not exceed 6 m.