Hold value elimination – KEYENCE LS-7000 User Manual
Page 104
4-52
Chapter 4 Function Settings
4
1
Press the key corresponding to the
type of tolerance to be set.
The corresponding tolerance item
flashes.
The following display flashes.
Value
Tolerance display
HI set value
HH set value
LO set value
LL set value
HI
GO
HI
GO
GO
LO
GO
LO
HI
GO
LO
2
Press the
or
key and set the
tolerance.
3
Press the flashing tolerance key
again.
The input value is entered.
PROG No.
AREA
DISP
ESC
G
C
ENT
HOLD
ZERO
FOCUS
A
POSITION
HI
GO
LO
Press these keys to set the value.
Hold Value Elimination
If the target has dirt or dust, the measured value may be wide of the accurate value. The
LS-7000 Series automatically detects and eliminates such values.
Hold Value Elimination Conditions
Hold values are eliminated under the following conditions.
● In Tolerance Mode with Reference Value Used
• The measured value is below the reference value + lower limit hold value (i.e., Measured
value < (Reference value + Lower limit hold value)
• The measured value is above the reference value + upper limit hold value (i.e., Refer-
ence value + Upper limit hold value) < Measured value
● In Tolerance Mode with Threshold Used
• The measured value is below the lower limit hold value (i.e., Measured value < Lower
limit hold value)
• The measured value is above the upper limit hold value (i.e., Upper limit hold value <
Measured value)
Note:
• Hold values are eliminated before measured values are averaged.
• The previous measured value will be kept on hold and displayed when the
present hold value is eliminated.