beautypg.com

Rockwell Automation 441-PEN25 Enpac Ex Data Collector User Manual

Page 93

background image

Publication GMSI00-UM001A-EN-E - February 2005

Collecting and Reviewing Data 87

3. If you set Auto Advance to "On" in the Instrument Configuration

screen, the Enpac Ex automatically goes to the next point in the list
after if finishes collecting the time waveform measurement. If you set
Auto Advance to "Off" then the Time Waveform Measurement screen
appears.

4. Use the LEFT and RIGHT ARROW keys to move the frequency cursor

Use the UP and DOWN ARROW keys to move the amplitude cursor.

5. Press the +/- key to expand and compress the time waveform spectrum

plot.

6. When finished viewing the time waveform, press READ/OK to

continue to the next point.

Collecting Magnitude and Phase Measurements at Orders

The Enpac Ex can collect multiple magnitude and phase at orders of the
operating speed. You use a tachometer signal measuring the operating speed to
collect phase data.

Follow these steps to collect a spectrum for calculating additional magnitude
and phase values at orders of the operating speed.

1. Align the speed reference device, such as a tachometer, in the correct

direction and attach the vibration transducer properly.

TIP

The green LED illuminates when the measurement is
complete. See LED Indicators on page 12.

WARNING

Do not connect the tachometer to the Enpac Ex unless the
area is known to be nonhazardous.