Enpac Ex Parameters

Appendix B provides a list and description of the Enpac Ex parameters.

Instrument Setup
Parameters

The Instrument Setup parameters define the global options for the Enpac Ex
including the date, time, communication baud rate, and time-out. Use the
Touchscreen parameter to re-calibrate the touchscreen. The settings are
stored in Flash memory and will be reset to the factory defaults on a hardware
reset.

For information about

See page

Instrument Setup Parameters

105

Instrument Configuration Parameters

107

Offroute User Parameters

112

Instrument Setup Parameters

Parameter Name

Description

Values/Comments

Timeout

The number of minutes the Enpac Ex can remain idle
before it shuts itself off to conserve battery power.

Press the ON/OFF key to reactivate the instrument
following a timeout. The instrument returns to the
point in the program where it was when it shut off.
No stored data will be lost.

Options: 5 mins

10 mins
15 mins
20 mins
Off (default)

IrDA

Enables or disables the Infra-Red serial port.

The Enpac Ex does not currently
support this interface. We recommend
you set this option to "Disabled"
(default).

Baud Rate

The baud rate for the serial communication.

Options: 1200

2400
4800
9600
19200 (default)
38400
115200

Dflt Units

Defines the units of measurement for off route
standard velocity measurements.

Options: English (ips) (default)

Metric (mm/s)

Important: Used only for Off route
standard velocity measurements.

Date Format

The format of the date displayed in the Enpac Ex.

Options: DD/MM/YYYY

MM/DD/YYYY (default)
YYYY/MM/DD