background image

COMMAND DETAILS

95

:TIMebase:WINDow:SCALe

Description

Sets or queries the zoom horizontal scale.

Note

If the oscilloscope is under “ZOOM” mode, the
main timebase function will be disabled and
cannot be modified.

Syntax

:TIMebase:WINDow:SCALe { | ?}

Related
commands

:TIMebase:MODe

Parameter

Zoom horizontal scale. The range will
depend on the time base.

Return parameter

Returns the zoom horizontal scale.

Example

:TIMebase:WINDow:SCALe 1.0E-3

Sets the zoom horizontal scale to 1ms.

:TIMebase:EXPand

Description

Sets or queries the horizontal expand mode.

Syntax

:TIMebase:EXPand { CENTer | TRIGger | ? }

Parameter

CENTer

Sets the horizontal expand mode by
center.

TRIGger

Sets the horizontal expand mode by
the trigger position.

Return parameter

Sets the horizontal expand mode (CENTER,
TRIGGER).

Example

:TIMebase:EXPand TRIGger

Sets the horizontal expansion from the trigger
position.

Set