beautypg.com

GW Instek GDS-200 Series Programming Manual User Manual

Page 26

background image

GDS-200 & GDS-300 Programming Manual

26

Example

*SAV 1
Saves the current setup to Set 1.


*SRE

Description

Sets or queries the Service Request Enable register.
The Service Request Enable register determines
which registers of the Status Byte register are able
to generate service requests.

Syntax

*SRE { | ?}

Parameter/
Return parameter

0~255

Bit Weight

Bit# Weight Event Description
0

1

Not used

1

2

Not used

2

4

Not used

3

8

Not used

4

16

MAV Message Available Bit

5

32

ESB

Event Status Bit

6

64

MSS

Master Summary Bit

6

64

RQS

Request Service Bit

7

128

Not used

Example

*SRE?
>48
Indicates that the MAVB and ESB bit are both set.

*STB

Description

Queries the bit sum of the Status Byte register with
MSS (Master summary Status).

Set

Query

Query

This manual is related to the following products: