COMMAND DETAILS

119

Parameter/

OFF

Pauses/Stops the test

Return Parameter ON

Starts the test

Example

:GONogo:EXECute OFF

Turns Go/NoGo off.

:GONogo:FUNCtion

Description

Initializes the Go/NoGo function. This must be
run before any Go/NoGo commands can be used.

Syntax

:GONogo:FUNCtion

:GONogo:NGCount

Description

Returns the Go/NoGo counter.

Syntax

:GONogo:NGCount?

Return parameter Returns a string in the following format “number of

violations,total tests”

Example

:GONogo:NGCount?
> 3,25
Indicates that 3 violations occurred over 25 tests.

:GONogo:NGDefine

Description

Sets the Go/NoGo “When” conditions.

Syntax

:GONogo:NGDefine {EXITs|ENTers|?}

Parameter/
Return Parameter

EXITs

Sets the NoGo condition to when the
input signal exceeds the limit boundary.

ENTers

Sets the NoGo condition to when the
input signal stays within the limit
boundary.

Example

:GONogo:NGDefine EXITs
Sets the Go/NoGo condition to EXITs.